Water Total Organic Carbon Test Machine Total Inorganic Carbon Test Equipment NPOC Measurement Machine
Application Scope
Widely used for the determination of total carbon (TC), total organic carbon (TOC), total inorganic carbon (TIC), non-purgeable total organic carbon (NPOC) and other parameters in surface water, groundwater, industrial water, domestic sewage, production wastewater, seawater, drinking water and its source water.
Instrument Features
- Adopts 24-bit A/D conversion technology and 32-bit signal processing technology.
- Automatic leak detection and control system: monitors gas flow in real time with an alarm function to prevent gas leakage.
- Four-in-one technology for TIC reaction cell: heating, cooling, dehydration and liquid level monitoring.
- Automatic micro-sampling with controllable sample injection volume.
- Supports online parameter setting, real-time monitoring, self-calibration and flow rate control.
- Real-time detection software: plots sample curves on the same coordinate for intuitive repeatability display; monitors instrument operation status at all times and compares measured parameters with set parameters.
- Micro-signal processing system features online setting, real-time monitoring, self-verification and flow rate control, ensuring superior instrument performance and safe operation.
- Automatic leak detection and control system: detects gas flow in real time with alarm settings to prevent gas leakage, maximizing the safety of operators and the instrument.
- Application of flow rate control signal processing technology shields the impact of flow fluctuations and ensures stable and accurate readings.
- Real-time detection software: draws sample curves under the same coordinate system to visually show repeatability; tracks instrument operating status continuously and compares actual measured parameters with preset values.
- Different control temperatures can be set according to the properties of various samples to achieve complete sample digestion and ensure more accurate and reliable measurement data.
- The instrument adopts an internal weak-current design for safe operation.
Water Total Organic Carbon Test Machine Total Inorganic Carbon Test Equipment NPOC Measurement Machine
Technical Specifications
| Item |
Specification |
| Model |
LS-TA-200 (Dry Method) |
| TOC Measurement Range |
0 ~ 30000 mg/L |
| Maximum Salt Tolerance |
85 g/L |
| TOC Measurement Items |
IC, TIC, TOC |
| Detector |
NDIR (Non-Dispersive Infrared Detector) |
| Operation Mode |
Computer-controlled |
| Application Sample |
Water samples |
| Gas Requirement |
Oxygen: ≥ 99.995% |
| Detection Limit |
50 μg/L |
| Digestion Principle |
High-temperature catalytic oxidation |
| Repeatability |
3% |
| Analysis Time |
1-4 minutes per parameter |
| Digestion Temperature |
0-1000 °C (adjustable) |
| Sample Temperature |
1 °C ~ 95 °C |
| Particle Size |
≤ 0.2 mm |
| Power Supply |
220 VAC, 50/60 Hz |
| Dimensions (LWH) |
430 * 455 * 440 mm |