I. Product Overview
The fully automatic multi-channel Vicat apparatus is primarily used to automatically measure the standard consistency, water requirement, and setting time of cement paste. It can also be used to automatically determine the initial and final setting times of gypsum.
This fully automatic multi-channel Vicat apparatus utilizes a fully automatic control program and can test up to eight samples. It comes with a built-in constant temperature and humidity curing chamber, allowing for real-time temperature and humidity control of the samples. The instrument automatically completes the timing, release, drop distance measurement, automatic needle removal, automatic cleaning, and test evaluation processes. The entire testing process complies with standard requirements and is easy to operate.
II. Implementation Standards:
GB/T 1346-2011 "Test Method for Standard Consistency Water Requirement, Setting Time, and Soundness of Cement"
GBT 17669.4-1999 "Construction Gypsum" Determination of Physical Properties of Pure Slurry
III. Features:
1. One instrument can measure up to 8 samples;
2. 7-inch color capacitive touchscreen display;
3. Chinese-language user interface;
4. Real-time date and time display;
5. Automatic constant temperature and humidity chamber with temperature setting and display function, 20±1°C; humidity setting and display function, ≥95%;
6. Automatic cleaning of the measuring probe to ensure measurement accuracy;
7. Timed screen saver function, energy-saving and environmentally friendly;
8. Automatic release function Drop assembly, free fall without manual intervention;
9. Automatically measures drop distance at a fixed time, eliminating the need for manual reading;
10. Automatically retrieves the dropped standard needle;
11. Equipped with a turntable for automatic rotation and adjustment of the test position;
12. Built-in large-capacity memory chip records all process data from 1,000 tests;
13. Automatically generates test record curves for easy observation of performance changes over time;
14. Automatic voice announcement function;
15. Equipped with a printer for direct printing of test results;
16. Equipped with a USB data output port.
Benchtop Cement Consistency Tester Automatic Cement Vicat Instrument Gypsum Setting Times Tester
Main Technical Specifications
|
1. Number of Samples Measured |
8 units |
|
2. Humidity Control |
≥95% |
|
3. Temperature Control |
20±1°C |
|
4. Temperature Display Range |
-55~125°C |
|
5. Temperature Resolution |
0.1°C |
|
6. Vertical Displacement Range |
0~100mm |
|
7. Displacement Resolution |
0.01mm |
|
8. Linear Displacement Measurement Accuracy |
≤75μm |
|
9. Automatic End Face Measurement |
Positioning accuracy ≤75μm |
|
10. Drop Distance Range |
0~100mm |
|
11. Drop Distance Resolution |
0.1mm |
|
12. Drop Release Time |
0~100s |
|
13. Initial Setting Time Accuracy |
±15 minutes |
|
14. Final Setting Time Accuracy |
±45 minutes |
|
15. Data Storage |
1000 sets |
|
16. Print Function |
Built-in micro-printer |
|
17. Standard Consistency Bar |
Effective length 50±1mm, diameter φ10±0.05mm |
|
18. Initial Setting Needle |
Effective length 50±1mm, diameter φ1.13±0.05mm |
|
19. Final Setting Needle |
Effective length 30±1mm, diameter φ1.13±0.05mm |
|
20. Total Mass of Standard Consistency Bar/Initial Setting Needle/Final Setting Needle + Drop Slide |
300±1g |
|
21. Test Mold |
Deep 40±0.2mm, φ65±0.5mm (top), φ75±0.5mm (bottom) |
|
22. Glass Plate |
100×100mm, thickness 4-5mm |
|
23. Power Supply Voltage |
AC 220V ±10%, 50Hz |
|
24. Power Supply Wattage |
1500W |
|
25. Operating Environment |
0-40°C, relative humidity not exceeding 85% |
|
26. Dimensions |
1000*850*1700mm |
|
27. Overall Weight |
Approximately 200kg |
Equipment Configuration List
|
1 |
Vicat instrument main unit |
|
1 |
|
2 |
Power cord |
3 cores |
1 |
|
3 |
Instruction manual |
|
1 |
|
4 |
Certificate of conformity |
|
1 |
|
5 |
Warranty card |
|
1 |
|
6 |
Vicat test mold |
|
1 |
|
7 |
Glass base plate |
|
1 |
|
8 |
Standard consistency rod |
|
1 |
|
9 |
Initial setting needle |
|
1 |
|
10 |
Final setting needle |
|
1 |