Customization: | Available |
---|---|
After-sales Service: | Online After-Sales Service |
Power Supply: | 220V |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Lab UV Aging Test Chamber Durable Environmental Climatic Test Chamber
Main technical parameters | |
Item | UV Lamp aging test chamber |
Model | LR-B002-B |
Interior size | W1150 X H400 x D500mm |
Exterior size | W1350 X H1450 x D650mm |
Weight | 120kgs |
Controller | LCD 7-inch touch screen controller, with programmable temp., humid., UV(sun), and time and Fix mode of 2 running modes. USB interface Operation button Alarm indicator Irradiation timer |
Temp. control mode | PID self-adjustment SSR control |
Temp. range | RT+10ºC ~ 70ºC (adjustable) |
Temp. deviation | ±2.0ºC |
Temp. fluctuation | ±0.5ºC |
Humidity range | ≥93%R.H |
Temp. uniformity | ±1ºC |
Temp. fluctuation | ±0.5ºC |
Distance between lamps | 75mm±15mm |
Distance between samples and lamps | 100 ~ 50mm±30mm |
Irradiation level | ≤1.1W/m2 adjustable |
Adjustment | Test period for Temp., irradiation, condensation, spray can be adjustable |
Irradiation lamp | ATLAS Lamp UVA-340, L=1200/40W, 8pcs (average service life: 1600hrs) |
Temp. control method | PID self-adjustable SSR control |
Depth of water channel | 25mm under auto-control |
UV Wavelength | UVA-340: 315 ~ 400nm; |
Test time | 0~999 H 99 min (adjustable) |
Function | Temperature, light, condensation, spray test cycle is adjustable |
Black board Temp. | 50ºC ~ 70ºC (temperature inside the chamber plus 20ºC ~ 25ºC) |
1.Structure & Material | |
Structure: 1.The middle of the machine is the working chamber, the water system located at the bottom of machine, its top is the control cabinet and control panel. Open the door from front direction. The bottom of the chamber with high-quality PU activity wheel that can be fixed, users can move it according to their requirements positioning. 2.Single cycling system for air flue in the box, there's one inlet axis weeding machine which improves the flow amount of air, the heating ability and the whole temperature uniformity in the test machine. Water source and consumption: Tap water or distilled water for 8liters per day. 3.Heating method is to heat liner water tank with fast heating speed and high temperature uniformity. Material: 4.Inner box: high grade stainless steel plate 5.Outer box: stainless steel plate SUS304 or stoving varnish for option. |