CRH-V Vacuum Glass Heat Transfer Coefficient Tester thermal conductivity meter
1.Product introduction
The instrument mainly measures the U(K) value of the heat transfer coefficient of vacuum glass, which complies with the standard JC/T2708-2022 "Test method for heat transfer coefficient of vacuum glass". It can also test homogeneous plate materials such as plastics, rubber, glass, fiber, benzene board, extruded board, foamed concrete, hollow glass, wood board, various insulation materials, etc. At the same time, it can measure the thermal conductivity of various materials such as granular materials, bulk materials, soft materials, etc. Applicable standards: GB/T10294-2008 "Determination of steady-state thermal resistance and related properties of insulation materials"; GB/T3399-1982 "Test method for thermal conductivity of plastics-heat-protected flat plate method"; GB/T10801.1-2002 "Polystyrene foam plastics for thermal insulation" specifies the measurement of thermal conductivity of polyurethane rigid foam materials; GB/T10801.2-2002 "Extruded polystyrene foam plastics for thermal insulation (XPS)" specifies the measurement of thermal conductivity of extruded polystyrene foam plastics; GB/T3139-2005 "Test method for thermal conductivity of fiber reinforced plastics"; GB/T 17794-2008 "Flexible foam rubber and plastic insulation products".
Product Parameters
CRH-V Vacuum Glass Heat Transfer Coefficient Tester thermal conductivity meter
2. Product Parameter
heat transfer coefficient U(k) measurement range
|
0.5 ~ 30 W/ .k
|
thermal conductivity measurement range
|
0.001 ~ 3.000 W/ m.K |
measurement accuracy
|
±3% |
Hot surface temperature measurement range
|
room temperature ~ 159ºC, temperature resolution: 0.01ºC |
Cold surface temperature measurement range
|
0 ~ 60ºC, temperature resolution: 0.01ºC |
pressure measurement range
|
0 ~ 1000N, resolution: 0.1N
|
Test piece size
|
300*300* (5 ~ 50) mm (basic configuration) (other specifications can be customized according to user requirements) |
power supply power |
220V 50HZ.500W |
full computer automatic detection and data processing system, compatible with the general computer.
3. complete set
1. One host computer;
2. A set of analysis software (Chinese and English versions);
3, HX-1005 precision low temperature constant temperature tank one (temperature control accuracy 0.01ºC);
4, a desktop Lenovo computer (users can choose);
5. A standard sample;
4. Dimensions and weight
1. Host: standard landing model: (length × width × height) 550×450×1500mm, net weight: 100Kg.
2, precision low temperature constant temperature tank: (length × width × height) 300×350×530mm, net weight: 35Kg.