Customization: | Available |
---|---|
After-sales Service: | Online |
Warranty: | 1 Year |
Still deciding? Get samples of US$ 12000/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
An overview
1. The whole system consists of P3000 high-pressure pump, UV3000 detector, sample valve, solvent tray, chromatographic column and working software.Installation, debugging, and usage are very simple.
2. The P3000 high pressure infusion pump is the core of CPLD technology, and the high precision infusion pump system is controlled by MCU.Full digital frequency system, advanced motor drive system and friendly man-machine display interface makes operation is very convenient, advanced pump head and single valve design train of thought, the experience of the most advanced in the world today, make the system under various conditions can ensure infusion precision and repeatability.
3. UV3000 is a kind of spectroscopic detector which can match any infusion pump, which can be used in chemical analysis, pesticide, pharmaceutical, chemical and other fields.The whole digital and high integration design concept is adopted to greatly reduce the noise of the instrument.Based on digital signal transmission and processing, the signal of the detector can be directly input to the serial port of the computer to collect, completely dumping the data acquisition card that the traditional instrument must use!
HPLC Liquid Chromatograph Tester for Chemical Analysis with High-Pressure Pump Chromatography Analyzer
Main features
1. P3000 series high voltage analysis/semi-preparation pumps
Dual plunger reciprocating pump, double plunger suction fluid.
Display 320X240 array large LCD display, blue backlight
System protection slow start, stop, maximum/minimum pressure adjustable, user data automatic storage
System connection 1/8 "pipe connection
Control the independent mode to set all working parameters through the front panel.Work in high
In the gradient mode, one of the pumps is the main pump, which controls the other types of high
Pressure infusion pump and the gradient curve that can be achieved by the screen
Verify.
The serial port mode is controlled by the RS232 port and the high-pressure infusion pump is controlled
Power supply 220V plus or minus 10%, 50Hz, 400W
2. UV3000UV/VIS detector
Control the independent mode via the front panel button
The serial port mode is controlled by RS232 software
Display 320X240 LCD, adjustable blue backlight
Power power 220V plus or minus 10%, 50Hz, 60W
Dimensions and weight: 398X149X267mm (deep X height X width), 14kg
GLP information general electric light time, electric light frequency, change of lamp number and product serial number, etc
HPLC Liquid Chromatograph Tester for Chemical Analysis with High-Pressure Pump Chromatography Analyzer
Technical parameters of P3000 high-pressure infusion pump:
Fluids Input:Double plunger reciprocating pump, a main suction plunger.One is for auxiliary plunger
Pump head:Analysis type: 10ml stainless steel/PEEK pump head/semi-preparation 50ml stainless steel/PEEK pump head
Maximum working pressure:Analysis type: 42MPa stainless steel pump head /30MPa PEEK pump head semi-preparation: 30MPa stainless steel pump head /20MPa PEEK pump
Flow range:Analysis type: 0.001-9.999ml semi-preparation type: 0.01-49.99 ml
Flow accuracy:RSD is less than 0.5% (1ml/min)
Flow repeatability:RSD less than 0.1% (1ml/min)
The display screen:320*240 array of LCD display, adjustable blue backlight.
System protection:Slow start, stop below minimum pressure 2ML.Maximum minimum pressure adjustable, user data automatic storage.
System connection:1/16 pipeline connection
Control system:Independent mode: full keyboard setting parameters.When work in high pressure gradient mode, one of the pump as the main pump, control of other same type high pressure infusion pump, and through the gradient curve to shown on the display of gradient method for validation.Serial port mode: through RS232, control the high pressure infusion pump.
Note: external dimensions: 398*149*267mm weight: 10 kg power supply: 220V plus or minus 10%, 50H Z, 40W
HPLC Liquid Chromatograph Tester for Chemical Analysis with High-Pressure Pump Chromatography Analyzer
Technical parameters of UV3000 detector
Wavelength range/accuracy:190-700nm + 2nm
Wavelength repeatability:0.4 nm
The light source: lamp (standard configuration);Tungsten lamp (replaceable)
Light source control:The lamp can be selected and the light is set
Scan function:The scan can be programmed to scan the starting point and ending point can be set. The scan step can be set.
Detection range:0-2 au
The baseline noise:1.5* 10-5au/h dynamic: velocity 1ml/min methanol/water = 80/20, wavelength 254nm
Baseline drift:4* 10-4au/h dynamic: velocity 1ml/min methanol/water = 80/20, wavelength 254nm
Time constant:0.5/0.1/0.2/1/2/5/10 s
Analog signal:Plus or minus 0.1V plus or minus 1.0v
Integral output:Adjustable level 16
Automatic zero:Full scale
Control:Independent mode: full keyboard setting parameters.Serial port mode: through RS232, control the high pressure infusion pump.
The display screen:320*240 array of LCD display, adjustable blue backlight.
GLP:Total lighting time, lighting times.Change the number of lights and serial number of products
Note: external dimensions: 398*149*267mm weight: 10 kg power supply: 220V plus or minus 10%, 50H Z, 40W