Customization: | Available |
---|---|
After-sales Service: | online |
Warranty: | 1 year |
Still deciding? Get samples of US$ 1400/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Freezing Strength Tester Freezing Force Tester
1. Model Advantages
1.1 The instrument comes standard with a 7-inch high-definition touch screen and interactive intelligent control.
1.2 The screen displays in real-time: Peak Bloomg, Final Bloomg, Average Bloomg, Penetration Depth mm, Standard Deviation, Cumulative Measurement Counts, and Hold Time Countdown.
1.3 The freezing strength resolution is 0.1 Bloomg.
1.4 Provides an option to output measurement data in Excel format via USB.
1.5 Measurement speed selection range: 0.1mm/s to 5mm/s with infinite adjustment, allowing for pre-measurement speed, during-measurement speed, and post-measurement speed settings.
1.6 High-speed operation with a noiseless mechanical structure.
1.7 Automatic zero calibration, with a standard gel strength measurement accuracy verification program.
Freezing Strength Tester Freezing Force Tester
1. Model Advantages
1.1 The instrument comes standard with a 7-inch high-definition touch screen and interactive intelligent control.
1.2 The screen displays in real-time: Peak Bloomg, Final Bloomg, Average Bloomg, Penetration Depth mm, Standard Deviation, Cumulative Measurement Counts, and Hold Time Countdown.
1.3 The freezing strength resolution is 0.1 Bloomg.
1.4 Provides an option to output measurement data in Excel format via USB.
1.5 Measurement speed selection range: 0.1mm/s to 5mm/s with infinite adjustment, allowing for pre-measurement speed, during-measurement speed, and post-measurement speed settings.
1.6 High-speed operation with a noiseless mechanical structure.
1.7 Automatic zero calibration, with a standard gel strength measurement accuracy verification program.
Freezing Strength Tester Freezing Force Tester
3.Features
2.1 Instrument Display Features
7-inch high-definition touch screen, interactive intelligent control.
2.2 System Access Management: Multi-level, multi-user (username + password login).
2.3 "Intelligent" Control: Simply press the "Start" button, and all parameters such as Peak Bloomg, Final Bloomg,Average Bloomg,Penetration Depth, Standard Deviation, and Cumulative Measurement Counts are displayed.
2.4 Comprehensive Measurement Parameters for Different Sample Properties: The system provides a full set of preset measurement parameters: trigger point, deformation, measurement speed, creep characteristic holding time, continuous cycle measurement counts, etc.
2.5 Protection Features
Upper and lower limit distance protection, force sensor overload protection, and measurement data protection.
2.6 Freezing Strength Measurement Accuracy Verification
The instrument comes with a verification operation interface; it offers a 5-continuous standard sample detection scheme, automatically reads and displays verification data, and directly determines whether it is qualified or out of specification. Verification results can be printed in real-time.
2.7 Automatic Correction of Freezing Strength Measurement Accuracy Deviation: The instrument has the ability to automatically correct the deviation of freezing strength measurement accuracy with a precision of 0.01g.
2.8 Deformation Displacement Accuracy Verification
The instrument comes with a 1~10mm displacement accuracy verification operation interface, automatically reads and displays verification data, and directly determines whether it is qualified or out of specification. Verification results can be printed in real-time.
2.9 High-Quality Vibration Damping Coupling Structure
The instrument is configured with a high-quality vibration damping coupling structure, ensuring no vibration or noise during measurement operation.
2.10 Quick Probe Change Interface
The instrument provides an interactive interface for quick probe changes, with automatic zero calibration after probe replacement.
4. Six Measurement Modes Provided by the Instrument
3.1 Bloom - Freezing Strength Test
Real-time peak load, peak deformation, final load, peak average, standard deviation, cumulative measurement counts.
3.2 Normal - Single Compression Test
Real-time load, positive peak, peak deformation, negative peak, final load, duration.
3.3 Hold Time - Compression Hold Test: Peak load, peak deformation, final load, real-time countdown.
3.4 Cycle Count - Multiple Compression Test
Maximum peak, minimum peak, peak average, last peak, standard deviation, remaining counts.
3.5 Freezing Strength Measurement Accuracy Verification
The instrument automatically measures accuracy, determines if it is out of specification, and corrects deviations if any.
3.6 Deformation Displacement Accuracy Verification
Can measure the positional error from 0 to 10mm at conventional measurement speeds and high-speed measurement speeds, determines if it is out of specification, automatically stores, and prints the verification results.
3.7 Measurement Units (Bloom g, gram, newton) One-Click Switch.
3.8 Output: RS232 Serial Print Interface. (Optional USB high-capacity access interface, output format
5.system configuration
Description |
Quantity |
JS-3C Freezing Strength Tester (2024 model with a measurement precision of 0.1g...) |
1 unit |
Cylindrical Probe: Diameter 12.700±0.010mm |
1 piece |
HW-3A Precision Constant Temperature Water Bath (Temperature range 3ºC~65ºC) (Food-grade thickened stainless steel lining) |
1 unit |
ZL-3 Refrigeration Unit (Can work continuously for over 72 hours) (Stainless steel evaporator) |
1 unit |
Freezing Strength Bottle (Capacity=150ml, Inner Diameter=59mm, Height=85mm) |
16 sets |
Rotatable and Telescopic Base Plate |
1 set |
232 Serial Print Interface |
Standard |
USB Output Interface |
Optional |
6. Technical Parameters
Description |
Specification |
Freezing Strength Measurement Range |
0~2000 Bloom g |
Freezing Strength Resolution |
0.1g |
Freezing Strength Measurement Accuracy |
≤±0.1% (Certificate of calibration with S/N provided) |
Repeated Measurement Accuracy |
≤±0.5 (Certificate of calibration with S/N provided) |
Deformation Displacement Range (Arbitrary Set Sampling Depth) |
0~75mm |
Displacement Resolution |
0.01mm (Real-time display) |
Conventional Measurement Speed Displacement Accuracy Error |
≤±0.05mm |
High-Speed Measurement Speed Displacement Accuracy Error |
≤±0.1mm |
Measurement Speed Selection Range |
0.1mm/s~5mm/s (Infinite adjustment for easier and more accurate measurement) |
Speed Resolution |
0.1mm/s |
Speed Travel Error |
≤0.1% (Set speed) |