Customization: | Available |
---|---|
After-sales Service: | 1 Year |
Warranty: | 1 Year |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
ASTM D2699 EN ISO 5165 Cetane Number and Octane Number Tester Testing Machine
Introduction
The A2020 octane number and cetane number analyzer is commonly used for on-site analysis of the octane number of automotive gasoline. It corresponds to the motor method and research method (RON and MON), and can also be applied to the analysis of cetane number of diesel.
Adaptation standards
ASTM D2699 EN ISO 5165
Instrument features
1.The analysis principle adopted is based on the insulation magnetic permeability of the octane value of gasoline and the cetane value of diesel, as well as the electrical charge characteristics of electromagnetic induction.
2. The instrument can measure minute changes in dielectric parameters and has an atmospheric pressure correction function.
3. Conduct more comprehensive and integrated measurements of various data related to petroleum products.
4. It can measure various types of gasoline containing additives.
5. Measure the cetane number, type and freezing point of the diesel.
6. At the same time, display RON, MON and the anti-explosion index (AKI). AKI = (RON + MON) / 2.
7. The powerful processing chip can quickly process data and is compatible with the WINDOWS system.
8. With temperature correction, the cost of use is low.
9. Simple and easy to operate, small in size, portable, with vibration-proof and solvent-proof features for the box, and sealed.
10. Four rows of backlit LCD display, suitable for low-temperature environments, power indicator, with an external low-voltage power supply.
ASTM D2699 EN ISO 5165 Cetane Number and Octane Number Tester Testing Machine
Product Parameter
Parameter Category |
Specific Parameters |
Octane number measurement range (ON, in ON units) |
40 - 120 |
Floatable range of octane number meter measurement results, per unit (maximum value) |
±0.2 |
Floatable range of hexadecane value meter measurement results: CN unit (maximum value) |
±0.5 |
Critical value of low battery voltage V |
5.4 |
Allowable measurement error of octane number meter |
Per unit (maximum value): 0.5 |
Allowable measurement error of hexadecane value meter, CN unit (maximum value) |
±1 |
Measurement time (seconds) |
1 - 5 |
Normal working time (unit: hour) |
1000 |
External dimensions |
Main unit: 100mm×210mm×40mm; Sensor: 60mm×100mm |
Instrument weight |
0.7kg |