Customization: | Available |
---|---|
Maxcapacity: | <1000KN |
Accuracy Grade: | 1 |
Still deciding? Get samples of US$ 3000/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Principle of instrumental measurement
Using radius measurement method, workpiece rotation. High precision air floating spindle is used as reference. Using imported precision circular grating, precision inductance displacement sensor and computer components. Measure the Angle and radial displacement of the measured workpiece to ensure the accuracy of the angular displacement and radial value of the measured workpiece; The measurement software adopts CA system measurement software based on Chinese Windows operating system platform to complete data collection, data processing and printing.
Cylindricity Measuring Instrument Cylindrical Measuring Device Concentricity Gauge Cylindricity Tester
Measurement functions
Measuring function: cylindricity meter, roundness, straightness, coaxiality, concentricity, end runout, end runout, verticality, parallelism, flatness
Analysis functions: surface waviness, spectrum analysis, wave height analysis, adjacent chip runout, segment difference
Roundness evaluation method (4 kinds) : minimum area method, least square method, minimum circumscribed circle method, maximum circumscribed circle
Roundness filtering method: 1, 1-500, 1-150, 1-50, 1-15, 15-500
Waviness filtering: 3-16, 3-17, 15-500, 17-100
Filtering form: Gaussian (ISO standard)
Cylindricity Measuring Instrument Cylindrical Measuring Device Concentricity Gauge Cylindricity Tester
Technical specification
Item |
Specification |
|
Measure range |
Max diameter |
Diameter 300mm |
Min inner diameter
|
Diameter 3mm |
|
Max height |
420mm |
|
Accuracy of principal axis |
radial error |
±(0.025+5H/10000)μm |
axial error |
±(0.025+6X/10000)μm |
|
Parallelism between axis of rotation and z-axis guide rail |
2μm/300mm(Busbar benchmark) |
|
workbench |
Mesa diameter |
Diameter 160mm |
swing diameter
|
Diameter 320mm |
|
load capacity |
30Kg |
|
adjustment range |
Adjustable eccentric±2mm ; Adjustable level±1° |
|
rotational speed
|
0-12 rpm(adjustable) |
|
Column |
Straightness (narrow range) |
0.5μm /100mm |
Straightness (full range) |
0.8μm /300mm |
|
horizontal arm
|
Horizontal travel distance |
160mm |
Move speed |
0.5-6mm/s OR MANUAL |
|
Sensor |
Stroke |
500μm (Radius difference) |
resolution
|
0.005μm |
|
Needle shape |
Ruby ball probe(diameter 2mm) |
|
Sensor direction/adjustment Angle |
either direction/±45° |
|
data collection |
Imports grating 7200point/week |
|
magnification |
Any to a maximum of 200,000 times |