Product introduction
This instrument adopts Karl - Fischer Coulometric titration to determine the trace moisture of liquid, gas and solid samples reliably.During the test, for the insoluble reagent solid and the substance that easily contaminates the electrode and reagent reaction, the corresponding solid, gas and liquid sampler can be used for indirect determination, which is a high-efficiency, full-automatic analytical instrument.Widely used in electric power, petroleum, chemical industry, medicine, railway, environmental protection, scientific research institutions and other industries.
Karl Fischer Oil Moisture Analyzer Titration Apparatus Coulometric Moisture Meter
Performance characteristics
1. High sensitivity: Karl fischer reagent and Coulomb titration were used for the test;
2. Automatic judgment of test end point, buzzer alarm prompt;
3. The dynamic curve of the test process shows the measured potential, and the test status is clear;
4. Automatically display, print and store test data;
5. Large data storage capacity, 1000 analysis and test records can be saved;
7. Adopt touch sliding agitator speed regulation;
8. There are 5 calculation formulas available.
Karl Fischer Oil Moisture Analyzer Titration Apparatus Coulometric Moisture Meter
Technical indicators
Display mode |
Touch high definition color LCD |
Status indication |
Display of the measured potential dynamic curve and written description |
Electrolytic current |
Maximum 400mA |
Measurement range |
0 g ~ 100mgH2O |
Resolution |
0.1 gH2O |
Accuracy |
0 g ~ 100ugH2O not greater than 3 g, greater than 100ugH2O not greater than 3 g±3% (excluding injection error) |
Data records |
1000 |
Printing |
Thermal printer |
Calculation formula |
5 kinds are optional |
Mixing speed regulation |
Panel touch sliding speed regulation |
Applicable ambient temperature |
5 ~ 35ºC |
Applicable relative humidity |
≤85% |
Power source |
AC220V±10%, 50Hz±5% |
Power rate |
30W |
Weight |
10kg |
Shape |
335*240*180mm |