Summary
The LT2800 is a new generation of intelligent laser particle size analysis system developed based on years of scientific research, with several innovative and patented technologies, and no need to change lenses or use standard samples for calibration. The dual-light source system with a wavelength of 638nm and a maximum power of 20mW solid laser and a wavelength of 468nm and a power of 12mW solid blue light source has a measuring range of 0.01 micron to 2800 micron and a measuring speed of up to 20,000 times per second, with extremely high sensitivity and reproducibility. Accurate and reliable results can be obtained for all types of samples, and an improved inversion algorithm is used to simulate the results of different instruments and methods.
LT2800 Scattering Particle Analyzer ISO13320 Particle Size Analysis Machine
Technical characteristics
1. The instrument fully complies with ISO13320 particle size measurement technical standards
2. Particle size range 0.01 micron - 2800 micron, single lens and no need to replace
3. Particle size measurement accuracy is better than ±0.5%, repeatability is better than ±0.5%
4. Integrated constant temperature system wavelength 638nm, the maximum power of 20mW solid laser and wavelength 468nm, the power of 12mW solid blue light source dual light source system, completely overcome the helium neon gas laser preheating time is long, short service life shortcomings
5. High sensitivity light energy tracking and stabilization technology
6. The detector is equipped with non-uniform cross area compensation detector array and high sensitivity backscatter detection unit
7. Optical fiber polarization space filter * with patented technology
8. Measuring speed up to 20000 times per second
9. Adopt oblique measurement window to reduce the influence of total reflection on the result
10. Intelligent automatic optical alignment
11. Automatic correction function of anomalous change of diffraction airy spot
12. Improved inversion algorithm, with the result simulation function of different instruments and methods
13. Equipped with independent automatic sampling system, and with SOP control and sharing function
14. Dual drive injection decentralized integration technology, circulation and mixing separate independent setting and control
15. Liquid level automatic induction and continuous adjustable control, the volume of dispersed medium can be controlled by software adjustment
16. The ultrasonic power of the sampler is 100 W, the ultrasonic power is continuously adjustable, and it has the function of low liquid level ultrasonic protection
LT2800 Scattering Particle Analyzer ISO13320 Particle Size Analysis Machine
Parameter
principle of measurement | Full range laser diffraction |
optical model | Complete Michelian theory and Fraunhofer theory are optional |
particle size range | 0.01μm -2800μm, no need to replace the lens |
detecting system | It includes a non-uniform cross area compensation detector array and a large Angle and backscatter detection system |
sampling method | A 638nm, 20mW solid laser with integrated thermostatic system and a 468nm solid blue light source with a maximum of 12mW |
light source | 638nm solid-state laser with integrated thermostatic system, up to 20mW |
Spatial filtering mode | Non-pinhole polarization filter patented technology |
measuring cell | Parallel slant |
Optical alignment system | Intelligent fully automatic |
measuring speed | Up to 20,000 times per second |
measuring time | Typical values are less than 10 seconds |
degree of accuracy | Dv50 better than ±0.6% (NIST-Traceable latex standard) |
repeatability | Dv50 better than ±0.5% (NIST-Traceable latex standard) |
laser safety | Class 1 laser products |
computer configuration | Intel i5 processor, 4GB RAM, 250GB hard drive, mouse, keyboard and widescreen display |
computer interface | USB2.0 or above |
Software running platform | Windows7 or above Professional Edition |
operating ambient temperature | 5ºC-40ºC |
Operating ambient humidity | 10%-85% relative humidity (no condensation) |
power requirement | 110V-240V, 50HZ-60Hz, standard grounding |
Optical system weight | 40KG |